Part II Volumetric Analysis
The processes of volumetric analysis are, in general, simpler than those of gravimetric analysis and accordingly serve best as an introduction to the practice of quantitative analysis. For their execution there are required, first, an accurate balance with which to weigh the material for analysis; second, graduated instruments in which to measure the volume of the solutions employed; third, standard solutions, that is, solutions the value of which is accurately known; and fourth, indicators, which will furnish accurate evidence of the point at which the desired reaction is completed. The nature of the indicators employed will be explained in connection with the different analyses.
The process whereby a!standard solution! is brought into reaction is called!titration!, and the point at which the reaction is exactly completed is called the!end-point!. The!indicator! should show the!end-point! of the!titration!. The volume of the standard solution used then furnishes the measure of the substance to be determined as truly as if that substance had been separated and weighed.
The processes of volumetric analysis are easily classified, according to their character, into:
I. NEUTRALIZATION METHODS; such, for example, as those of acidimetry and alkalimetry.
II. OXIDATION PROCESSES; as exemplified in the determination of ferrous iron by its oxidation with potassium bichromate.
III. PRECIPITATION METHODS; of which the titration for silver with potassium thiocyanate solution is an illustration.
From a somewhat different standpoint the methods in each case may be subdivided into (a) DIRECT METHODS, in which the substance to be measured is directly determined by titration to an end-point with a standard solution; and (b) INDIRECT METHODS, in which the substance itself is not measured, but a quantity of reagent is added which is known to be an excess with respect to a specific reaction, and the unused excess determined by titration. Examples of the latter class will be pointed out as they occur in the procedures.
Measuring Instruments
The Analytical Balance
For a complete discussion of the physical principles underlying the construction and use of balances, and the various methods of weighing, the student is referred to larger manuals of Quantitative Analysis, such as those of Fresenius, or Treadwell-Hall, and particularly to the admirable discussion of this topic in Morse’s!Exercises in Quantitative Chemistry!.
The statements and rules of procedure which follow are sufficient for the intelligent use of an analytical balance in connection with processes prescribed in this introductory manual. It is, however, imperative that the student should make himself familiar with these essential features of the balance, and its use. He should fully realize that the analytical balance is a delicate instrument which will render excellent service under careful treatment, but such treatment is an essential condition if its accuracy is to be depended upon. He should also understand that no set of rules, however complete, can do away with the necessity for a sense of personal responsibility, since by carelessness he can render inaccurate not only his own analyses, but those of all other students using the same balance.
Before making any weighings the student should seat himself before a balance and observe the following details of construction:
1. The balance case is mounted on three brass legs, which should preferably rest in glass cups, backed with rubber to prevent slipping. The front legs are adjustable as to height and are used to level the balance case; the rear leg is of permanent length.
2. The front of the case may be raised to give access to the balance. In some makes doors are provided also at the ends of the balance case.
3. The balance beam is mounted upon an upright in the center of the case on the top of which is an inlaid agate plate. To the center of the beam there is attached a steel or agate knife-edge on which the beam oscillates when it rests on the agate plate.
4. The balance beam, extending to the right and left, is graduated along its upper edge, usually on both sides, and has at its extremities two agate or steel knife-edges from which are suspended stirrups. Each of these stirrups has an agate plate which, when the balance is in action, rests upon the corresponding knife-edge of the beam. The balance pans are suspended from the stirrups.
5. A pointer is attached to the center of the beam, and as the beam oscillates this pointer moves in front of a scale near the base of the post.
6. At the base of the post, usually in the rear, is a spirit-level.
7. Within the upright is a mechanism, controlled by a knob at the front of the balance case, which is so arranged as to raise the entire beam slightly above the level at which the knife-edges are in contact with the agate plates. When the balance is not in use the beam must be supported by this device since, otherwise, the constant jarring to which a balance is inevitably subjected, will soon dull the knife-edges, and lessen the sensitiveness of the balance.
8. A small weight, or bob, is attached to the pointer (or sometimes to the beam) by which the center of gravity of the beam and its attachments may be regulated. The center of gravity must lie very slightly below the level of the agate plates to secure the desired sensitiveness of the balance. This is provided for when the balance is set up and very rarely requires alteration. The student should never attempt to change this adjustment.
9. Below the balance pans are two pan-arrests operated by a button from the front of the case. These arrests exert a very slight upward pressure upon the pans and minimize the displacement of the beam when objects or weights are being placed upon the pans.
10. A movable rod, operated from one end of the balance case, extends over the balance beam and carries a small wire weight, called a rider. By means of this rod the rider can be placed upon any desired division of the scale on the balance beam. Each numbered division on the beam corresponds to one milligram, and the use of the rider obviates the placing of very small fractional weights on the balance pan.
If a new rider is purchased, or an old one replaced, care must be taken that its weight corresponds to the graduations on the beam of the balance on which it is to be used. The weight of the rider in milligrams must be equal to the number of large divisions (5, 6, 10, or 12) between the central knife-edge and the knife-edge at the end of the beam. It should be noted that on some balances the last division bears no number. Each new rider should be tested against a 5 or 10-milligram weight.
In some of the most recent forms of the balance a chain device replaces the smaller weights and the use of the rider as just described.
Before using a balance, it is always best to test its adjustment. This is absolutely necessary if the balance is used by several workers; it is always a wise precaution under any conditions. For this purpose, brush off the balance pans with a soft camel’s hair brush. Then note (1) whether the balance is level; (2) that the mechanism for raising and lowering the beams works smoothly; (3) that the pan-arrests touch the pans when the beam is lowered; and (4) that the needle swings equal distances on either side of the zero-point when set in motion without any load on the pans. If the latter condition is not fulfilled, the balance should be adjusted by means of the adjusting screw at the end of the beam unless the variation is not more than one division on the scale; it is often better to make a proper allowance for this small zero error than to disturb the balance by an attempt at correction. Unless a student thoroughly understands the construction of a balance he should never attempt to make adjustments, but should apply to the instructor in charge.
The object to be weighed should be placed on the left-hand balance pan and the weights upon the right-hand pan. Every substance which could attack the metal of the balance pan should be weighed upon a watch-glass, and all objects must be dry and cold. A warm body gives rise to air currents which vitiate the accuracy of the weighing.
The weights should be applied in the order in which they occur in the weight-box (not at haphazard), beginning with the largest weight which is apparently required. After a weight has been placed upon the pan the beam should be lowered upon its knife-edges, and, if necessary, the pan-arrests depressed. The movement of the pointer will then indicate whether the weight applied is too great or too small. When the weight has been ascertained, by the successive addition of small weights, to the nearest 5 or 10 milligrams, the weighing is completed by the use of the rider. The correct weight is that which causes the pointer to swing an equal number of divisions to the right and left of the zero-point, when the pointer traverses not less than five divisions on either side.
The balance case should always be closed during the final weighing, while the rider is being used, to protect the pans from the effect of air currents.
Before the final determination of an exact weight the beam should always be lifted from the knife-edges and again lowered into place, as it frequently happens that the scale pans are, in spite of the pan-arrests, slightly twisted by the impact of the weights, the beam being thereby virtually lengthened or shortened. Lifting the beam restores the proper alignment.
The beam should never be set in motion by lowering it forcibly upon the knife-edges, nor by touching the pans, but rather by lifting the rider (unless the balance be provided with some of the newer devices for the purpose), and the swing should be arrested only when the needle approaches zero on the scale, otherwise the knife-edges become dull. For the same reason the beam should never be left upon its knife-edges, nor should weights be removed from or placed on the pans without supporting the beam, except in the case of the small fractional weights.
When the process of weighing has been completed, the weight should be recorded in the notebook by first noting the vacant spaces in the weight-box, and then checking the weight by again noting the weights as they are removed from the pan. This practice will often detect and avoid errors. It is obvious that the weights should always be returned to their proper places in the box, and be handled only with pincers.
It should be borne in mind that if the mechanism of a balance is deranged or if any substance is spilled upon the pans or in the balance case, the damage should be reported at once. In many instances serious harm can be averted by prompt action when delay might ruin the balance.
Samples for analysis are commonly weighed in small tubes with cork stoppers. Since the stoppers are likely to change in weight from the varying amounts of moisture absorbed from the atmosphere, it is necessary to confirm the recorded weight of a tube which has been unused for some time before weighing out a new portion of substance from it.
Weights
The sets of weights commonly used in analytical chemistry range from 20 grams to 5 milligrams. The weights from 20 grams to 1 gram are usually of brass, lacquered or gold plated. The fractional weights are of German silver, gold, platinum or aluminium. The rider is of platinum or aluminium wire.
The sets of weights purchased from reputable dealers are usually sufficiently accurate for analytical work. It is not necessary that such a set should be strictly exact in comparison with the absolute standard of weight, provided they are relatively correct among themselves, and provided the same set of weights is used in all weighings made during a given analysis. The analyst should assure himself that the weights in a set previously unfamiliar to him are relatively correct by a few simple tests. For example, he should make sure that in his set two weights of the same denomination (i.e., two 10-gram weights, or the two 100-milligram weights) are actually equal and interchangeable, or that the 500-milligram weight is equal to the sum of the 200, 100, 100, 50, 20, 20 and 10-milligram weights combined, and so on. If discrepancies of more than a few tenths of a milligram (depending upon the total weight involved) are found, the weights should be returned for correction. The rider should also be compared with a 5 or 10-milligram weight.
In an instructional laboratory appreciable errors should be reported to the instructor in charge for his consideration.
When the highest accuracy is desired, the weights may be calibrated and corrections applied. A calibration procedure is described in a paper by T.W. Richards,!J. Am. Chem. Soc.!, 22, 144, and in many large text-books.
Weights are inevitably subject to corrosion if not properly protected at all times, and are liable to damage unless handled with great care. It is obvious that anything which alters the weight of a single piece in an analytical set will introduce an error in every weighing made in which that piece is used. This source of error is often extremely obscure and difficult to detect. The only safeguard against such errors is to be found in scrupulous care in handling and protection on the part of the analyst, and an equal insistence that if several analysts use the same set of weights, each shall realize his responsibility for the work of others as well as his own.
Burettes
A burette is made from a glass tube which is as uniformly cylindrical as possible, and of such a bore that the divisions which are etched upon its surface shall correspond closely to actual contents.
The tube is contracted at one extremity, and terminates in either a glass stopcock and delivery-tube, or in such a manner that a piece of rubber tubing may be firmly attached, connecting a delivery-tube of glass. The rubber tubing is closed by means of a glass bead. Burettes of the latter type will be referred to as “plain burettes.”
The graduations are usually numbered in cubic centimeters, and the latter are subdivided into tenths.
One burette of each type is desirable for the analytical procedures which follow.
Preparation of a Burette for Use
The inner surface of a burette must be thoroughly cleaned in order that the liquid as drawn out may drain away completely, without leaving drops upon the sides. This is best accomplished by treating the inside of the burette with a warm solution of chromic acid in concentrated sulphuric acid, applied as follows: If the burette is of the “plain” type, first remove the rubber tip and force the lower end of the burette into a medium-sized cork stopper. Nearly fill the burette with the chromic acid solution, close the upper end with a cork stopper and tip the burette backward and forward in such a way as to bring the solution into contact with the entire inner surface. Remove the stopper and pour the solution into a stock bottle to be kept for further use, and rinse out the burette with water several times. Unless the water then runs freely from the burette without leaving drops adhering to the sides, the process must be repeated (Note 1).
If the burette has a glass stopcock, this should be removed after the cleaning and wiped, and also the inside of the ground joint. The surface of the stopcock should then be smeared with a thin coating of vaseline and replaced. It should be attached to the burette by means of a wire, or elastic band, to lessen the danger of breakage.
Fill the burettes with distilled water, and allow the water to run out through the stopcock or rubber tip until convinced that no air bubbles are inclosed (Note 2). Fill the burette to a point above the zero-point and draw off the water until the meniscus is just below that mark. It is then ready for calibration.
Reading of a Burette
All liquids when placed in a burette form what is called a meniscus at their upper surfaces. In the case of liquids such as water or aqueous solutions this meniscus is concave, and when the liquids are transparent accurate readings are best obtained by observing the position on the graduated scales of the lowest point of the meniscus. This can best be done as follows: Wrap around the burette a piece of colored paper, the straight, smooth edges of which are held evenly together with the colored side next to the burette (Note 1). Hold the paper about two small divisions below the meniscus and raise or lower the level of the eyes until the edge of the paper at the back of the burette is just hidden from the eye by that in front (Note 2). Note the position of the lowest point of the curve of the meniscus, estimating the tenths of the small divisions, thus reading its position to hundredths of a cubic centimeter.
Calibration of Glass Measuring Devices
If accuracy of results is to be attained, the correctness of all measuring instruments must be tested. None of the apparatus offered for sale can be implicitly relied upon except those more expensive instruments which are accompanied by a certificate from the!National Bureau of Standards! at Washington, or other equally authentic source.
The bore of burettes is subject to accidental variations, and since the graduations are applied by machine without regard to such variations of bore, local errors result.
The process of testing these instruments is called!calibration!. It is usually accomplished by comparing the actual weight of water contained in the instrument with its apparent volume.
There is, unfortunately, no uniform standard of volume which has been adopted for general use in all laboratories. It has been variously proposed to consider the volume of 1000 grams of water at 4°, 15.5°, 16°, 17.5°, and even 20°C., as a liter for practical purposes, and to consider the cubic centimeter to be one one-thousandth of that volume. The true liter is the volume of 1000 grams of water at 4°C.; but this is obviously a lower temperature than that commonly found in laboratories, and involves the constant use of corrections if taken as a laboratory standard. Many laboratories use 15.5°C. (60° F.) as the working standard. It is plain that any temperature which is deemed most convenient might be chosen for a particular laboratory, but it cannot be too strongly emphasized that all measuring instruments, including burettes, pipettes, and flasks, should be calibrated at that temperature in order that the contents of each burette, pipette, etc., shall be comparable with that of every other instrument, thus permitting general interchange and substitution. For example, it is obvious that if it is desired to remove exactly 50 cc. from a solution which has been diluted to 500 cc. in a graduated flask, the 50 cc. flask or pipette used to remove the fractional portion must give a correct reading at the same temperature as the 500 cc. flask. Similarly, a burette used for the titration of the 50 cc. of solution removed should be calibrated under the same conditions as the measuring flasks or pipettes employed with it.
The student should also keep constantly in mind the fact that all volumetric operations, to be exact, should be carried out as nearly at a constant temperature as is practicable. The spot selected for such work should therefore be subject to a minimum of temperature variations, and should have as nearly the average temperature of the laboratory as is possible. In all work, whether of calibration, standardization, or analysis, the temperature of the liquids employed must be taken into account, and if the temperature of these liquids varies more than 3° or 4° from the standard temperature chosen for the laboratory, corrections must be applied for errors due to expansion or contraction, since volumes of a liquid measured at different times are comparable only under like conditions as to temperature. Data to be used for this purpose are given in the Appendix. Neglect of this correction is frequently an avoidable source of error and annoyance in otherwise excellent work. The temperature of all solutions at the time of standardization should be recorded to facilitate the application of temperature corrections, if such are necessary at any later time.
Calibration of the Burettes
Two burettes, one at least of which should have a glass stopper, are required throughout the volumetric work. Both burettes should be calibrated by the student to whom they are assigned.
PROCEDURE.—Weigh a 50 cc., flat-bottomed flask (preferably a light-weight flask), which must be dry on the outside, to the nearest centigram. Record the weight in the notebook. (See Appendix for suggestions as to records.) Place the flask under the burette and draw out into it about 10 cc. of water, removing any drop on the tip by touching it against the inside of the neck of the flask. Do not attempt to stop exactly at the 10 cc. mark, but do not vary more than 0.1 cc. from it. Note the time, and at the expiration of three minutes (or longer) read the burette accurately, and record the reading in the notebook (Note 1). Meanwhile weigh the flask and water to centigrams and record its weight (Note 2). Draw off the liquid from 10 cc. to about 20 cc. into the same flask without emptying it; weigh, and at the expiration of three minutes take the reading, and so on throughout the length of the burette. When it is completed, refill the burette and check the first calibration.
The differences in readings represent the apparent volumes, the differences in weights the true volumes. For example, if an apparent volume of 10.05 cc. is found to weigh 10.03 grams, it may be assumed with sufficient accuracy that the error in that 10 cc. amounts to -0.02 cc., or -0.002 for each cubic centimeter (Note 3).
In the calculation of corrections the temperature of the water must be taken into account, if this varies more than 4°C. from the laboratory standard temperature, consulting the table of densities of water in the Appendix.
From the final data, plot the corrections to be applied so that they may be easily read for each cubic centimeter throughout the burette. The total correction at each 10 cc. may also be written on the burette with a diamond, or etching ink, for permanence of record.
Pipettes
A!pipette! may consist of a narrow tube, in the middle of which is blown a bulb of a capacity a little less than that which it is desired to measure by the pipette; or it may be a miniature burette, without the stopcock or rubber tip at the lower extremity. In either case, the flow of liquid is regulated by the pressure of the finger on the top, which governs the admission of the air.
Pipettes are usually already graduated when purchased, but they require calibration for accurate work.
Calibration of Pipettes
PROCEDURE.—Clean the pipette. Draw distilled water into it by sucking at the upper end until the water is well above the graduation mark. Quickly place the forefinger over the top of the tube, thus preventing the entrance of air and holding the water in the pipette. Cautiously admit a little air by releasing the pressure of the finger, and allow the level of the water to fall until the lowest point of the meniscus is level with the graduation. Hold the water at that point by pressure of the finger and then allow the water to run out from the pipette into a small tared, or weighed, beaker or flask. After a definite time interval, usually two to three minutes, touch the end of the pipette against the side of the beaker or flask to remove any liquid adhering to it (Note 1). The increase in weight of the flask in grams represents the volume of the water in cubic centimeters delivered by the pipette. Calculate the necessary correction.
Flasks
!Graduated or measuring flasks! are similar to the ordinary flat-bottomed flasks, but are provided with long, narrow necks in order that slight variations in the position of the meniscus with respect to the graduation shall represent a minimum volume of liquid. The flasks must be of such a capacity that, when filled with the specified volume, the liquid rises well into the neck.
Graduation of Flasks
It is a general custom to purchase the flasks ungraduated and to graduate them for use under standard conditions selected for the laboratory in question. They may be graduated for “contents” or “delivery.” When graduated for “contents” they contain a specified volume when filled to the graduation at a specified temperature, and require to be washed out in order to remove all of the solution from the flask. Flasks graduated for “delivery” will deliver the specified volume of a liquid without rinsing. A flask may, of course, be graduated for both contents and delivery by placing two graduation marks upon it.
PROCEDURE.—To calibrate a flask for!contents!, proceed as follows: Clean the flask, using a chromic acid solution, and dry it carefully outside and inside. Tare it accurately; pour water into the flask until the weight of the latter counterbalances weights on the opposite pan which equal in grams the number of cubic centimeters of water which the flask is to contain. Remove any excess of water with the aid of filter paper (Note 1). Take the flask from the balance, stopper it, place it in a bath at the desired temperature, usually 15.5° or 17.5°C., and after an hour mark on the neck with a diamond the location of the lowest point of the meniscus (Note 2). The mark may be etched upon the flask by hydrofluoric acid, or by the use of an etching ink now commonly sold on the market.
To graduate a flask which is designed to!deliver! a specified volume, proceed as follows: Clean the flask as usual and wipe all moisture from the outside. Fill it with distilled water. Pour out the water and allow the water to drain from the flask for three minutes. Counterbalance the flask with weights to the nearest centigram. Add weights corresponding in grams to the volume desired, and add distilled water to counterbalance these weights. An excess of water, or water adhering to the neck of the flask, may be removed by means of a strip of clean filter paper. Stopper the flask, place it in a bath at 15.5°C. or 17.5°C. and, after an hour, mark the location of the lowest point of the meniscus, as described above.
Notes
1 The inner surface of the burette must be absolutely clean if the liquid is to run off freely. Chromic acid in sulphuric acid is usually found to be the best cleansing agent, but the mixture must be warm and concentrated. The solution can be prepared by pouring over a few crystals of potassium bichromate a little water and then adding concentrated sulphuric acid.
2 It is always necessary to insure the absence of air bubbles in the tips or stopcocks. The treatment described above will usually accomplish this, but, in the case of plain burettes it is sometimes better to allow a little of the liquid to flow out of the tip while it is bent upwards. Any air which may be entrapped then rises with the liquid and escapes.
If air bubbles escape during subsequent calibration or titration, an error is introduced which vitiates the results.
1 The ends of the colored paper used as an aid to accurate readings may be fastened together by means of a gummed label. The paper may then remain on the burette and be ready for immediate use by sliding it up or down, as required.
2 To obtain an accurate reading the eye must be very nearly on a level with the meniscus. This is secured by the use of the paper as described. The student should observe by trial how a reading is affected when the meniscus is viewed from above or below.
The eye soon becomes accustomed to estimating the tenths of the divisions. If the paper is held as directed, two divisions below the meniscus, one whole division is visible to correct the judgment. It is not well to attempt to bring the meniscus exactly to a division mark on the burette. Such readings are usually less accurate than those in which the tenths of a division are estimated.
1 A small quantity of liquid at first adheres to the side of even a clean burette. This slowly unites with the main body of liquid, but requires an appreciable time. Three minutes is a sufficient interval, but not too long, and should be adopted in every instance throughout the whole volumetric practice before final readings are recorded.
2 A comparatively rough balance, capable of weighing to centigrams, is sufficiently accurate for use in calibrations, for a moment’s reflection will show that it would be useless to weigh the water with an accuracy greater than that of the readings taken on the burette. The latter cannot exceed 0.01 cc. in accuracy, which corresponds to 0.01 gram.
The student should clearly understand that!all other weighings!, except those for calibration, should be made accurately to 0.0001 gram, unless special directions are given to the contrary.
Corrections for temperature variations of less than 4°C. are negligible, as they amount to less than 0.01 gram for each 10 grams of water withdrawn.
3 Should the error discovered in any interval of 10 cc. on the burette exceed 0.10 cc., it is advisable to weigh small portions (even 1 cc.) to locate the position of the variation of bore in the tube rather than to distribute the correction uniformly over the corresponding 10 cc. The latter is the usual course for small corrections, and it is convenient to calculate the correction corresponding to each cubic centimeter and to record it in the form of a table or calibration card, or to plot a curve representing the values.
Burettes may also be calibrated by drawing off the liquid in successive portions through a 5 cc. pipette which has been accurately calibrated, as a substitute for weighing. If many burettes are to be tested, this is a more rapid method.
1 A definite interval must be allowed for draining, and a definite practice adopted with respect to the removal of the liquid which collects at the end of the tube, if the pipette is designed to deliver a specific volume when emptied. This liquid may be removed at the end of a definite interval either by touching the side of the vessel or by gently blowing out the last drops. Either practice, when adopted, must be uniformly adhered to.
1 The allowable error in counterbalancing the water and weights varies with the volume of the flask. It should not exceed one ten-thousandth of the weight of water.
2 Other methods are employed which involve the use of calibrated apparatus from which the desired volume of water may be run into the dry flask and the position of the meniscus marked directly upon it. For a description of a procedure which is most convenient when many flasks are to be calibrated, the student is referred to the!Am. Chem J.!, 16, 479.
Notes